Sign in →

Test Code 160 Uric Acid , Urine, 24 Hour

Additional Codes

COPIA Test Code EPIC Test Codes
Uric Acid, Urine 24hr LAB0200592 URIC ACID, URINE, 24 HR

 

Reporting Title

Uric Acid, Urine, 24hr

Methodology

Oxidation

Performing Laboratory

Rice Memorial Hospital

Specimen Requirements

Preferred Specimen Type: Urine, 24 hour

Container/Tube:

24 hour: Collection container (available from Rice Lab)
Random: Transport tube

Specimen Volume: 5 mL

Specimen Minimum Volume: 2 mL

Collection Instructions:

1. Add 10 mL 5% (12.5 N) NaOH to container before collection.

2. Instruct patient to void at the beginning of the collection period and discard the specimen.

3. Collect all urine, including the final specimen voided at the end of the collection period.

4. Measure and record urine volume, date and time collected, and collection interval.

5. Aliquot 5 mL of well mixed specimen into a pour off tube.

Additional Information: 

1. 24-Hour volume is required for timed urine collection.

2. Adjust pH of random urine with dropwise addition of 5% NaOH to acheive pH >8.

3. See Collection Instructions: Timed Urine Specimens in Special Instructions.

Specimen Transport Temperature

Preserved: Ambient up to 4 days

Reference Values

250-750 mg/24 hours

Day(s) Test Set Up

Monday through Sunday

CPT Coding

84560- Uric acid, urine

81050-Urine timed measurement (if appropriate)