Sign in →

Test Code LAB0249839 Rupture of Membranes

Additional Codes

EPIC Test Codes
LAB0249839 RUPTURE OF MEMBRANES

 

Reporting Title

Rupture of Membranes

Methodology

Immunochromatography

Performing Laboratory

Rice Memorial Hospital

Specimen Requirements

Specimen Type: Vaginal discharge

Container/Tube: Amnisure collection kit

Collection Instructions: 

1.  Collection (performed by Physician or RN on OB)

2.  With patient lying flat on her back, insert a sterile polyester swab (the polyester tip should not touch anything prior to insertion) 2-3 inches into the vagina for 1 minute to allow absorption of fluid into the swab.

3.  Place patient swab into vial and rinse in solvent by rotating for 1 minute.

4.  Remove swab and dispose of swab.

5.  Label blue capped solvent vial with patient information, record date /time of collection and collector.

6.  Place vial back into original plastic pouch. 

7.  Transport to Lab ASAP for testing. (Needs to arrive within 4 hours room temp (preferred) or < 6 hours refrigerated of collection for testing).

 

Specimen Transport Temperature

Analyze within 4 hours (preferred) or Refrigerated ≤6 hours

Reference Values

Negative if no rupture

Day(s) Test Set Up

Monday through Sunday

CPT Coding

84112